Method for measuring low activity extensive samples

This paper discusses a method of measuring low intensity gamma radiation from extensive sources using solid state detectors. In the first part the variation of the detector efficiency for a point source as a function of source position is measured. It is seen that the angular dependence is particula...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Tarela, Pablo Adrian, Mariscotti, Mario Alberto Juan, Perone, Elizabeth Andrea
Publicado: 1994
Acceso en línea:https://bibliotecadigital.exactas.uba.ar/collection/paper/document/paper_0168583X_v94_n4_p511_Tarela
http://hdl.handle.net/20.500.12110/paper_0168583X_v94_n4_p511_Tarela
Aporte de:
Descripción
Sumario:This paper discusses a method of measuring low intensity gamma radiation from extensive sources using solid state detectors. In the first part the variation of the detector efficiency for a point source as a function of source position is measured. It is seen that the angular dependence is particularly important. With this response function known, a semi empirical model is developed for the computation of the efficiency in cases of extensive radiation fields. Measurements are carried out which show the adequacy of the model. The method is suitable for liquid samples such as drinkable water, industrial waste liquids and sewage liquids. It is therefore useful for water quality monitoring and pollution control. © 1994.