Extension of the measurement range of MOS dosimeters using radiation induced charge neutralization
This work proposes a new measurement technique to extend the dose range of MOS dosimeters. The technique consists on alternating stages of positive oxide charge buildup with stages of radiation induced charge neutralization, maintaining a uniform sensitivity along the whole measurement. The techniqu...
Guardado en:
Publicado: |
2008
|
---|---|
Materias: | |
Acceso en línea: | https://bibliotecadigital.exactas.uba.ar/collection/paper/document/paper_00189499_v55_n4_p2141_Faigon http://hdl.handle.net/20.500.12110/paper_00189499_v55_n4_p2141_Faigon |
Aporte de: |
Sumario: | This work proposes a new measurement technique to extend the dose range of MOS dosimeters. The technique consists on alternating stages of positive oxide charge buildup with stages of radiation induced charge neutralization, maintaining a uniform sensitivity along the whole measurement. The technique was applied with 70 nm MOS dosimeters, extending the dose measurement range from less than a kilogray to more than 675 kGy without showing wear effects. An initial saturation of interface traps creation ensures repeatability in the responses. © 2008 IEEE. |
---|